ESSENTIAL SERIES

Valleylab Force 2 Electrosurgical Generator

The Valleylab Force 2 Electrosurgical Generator is a high-performance surgical device designed for precise cutting and coagulation in a wide range of procedures. It supports both monopolar and bipolar outputs, offering flexibility for different surgical techniques.

Built with advanced safety and control features, the system includes audible and visual alarms along with a patented Return Electrode Monitoring (REMβ„’) system that continuously checks patient impedance to prevent burns and ensure safe operation. Its versatile modes and stable power delivery make it suitable for general surgery, laparoscopic procedures, and other clinical applications requiring accuracy and reliability.

See full specs
Add benefits via tags or metafields
Tag products with feature:Title|Description or use the custom.benefits metafield for automatic feature generation.

The Valleylab Force 2 Electrosurgical Generator is a high-performance surgical device designed for precise cutting and coagulation in a wide range of procedures. It supports both monopolar and bipolar outputs, offering flexibility for different surgical techniques.

Built with advanced safety and control features, the system includes audible and visual alarms along with a patented Return Electrode Monitoring (REMβ„’) system that continuously checks patient impedance to prevent burns and ensure safe operation. Its versatile modes and stable power delivery make it suitable for general surgery, laparoscopic procedures, and other clinical applications requiring accuracy and reliability.

Product Specification
General Information
Device Type Electrosurgical Generator
Application Cutting and Coagulation
Output Modes Monopolar and Bipolar
Operating Modes
Cut Mode Standard cutting
Blend Modes Blend 1 / Blend 2 / Blend 3
Coagulation Standard Coagulation
Low Voltage Coag Controlled desiccation
Bipolar Mode Precision bipolar output
Power & Electrical
Maximum Output Power 300 W (Cut Mode)
Input Voltage 120 VAC (85–135 VAC range)
Frequency 45 – 65 Hz
Power Consumption 300 W
Output Waveforms
Cut 500 kHz sinusoidal waveform
Blend Modes 500 kHz burst sinusoidal (varying duty cycles)
Coagulation Damped sinusoidal burst
Low Voltage Coag Controlled burst waveform
Bipolar Unmodulated sinusoidal waveform
Output Characteristics
Cut Nominal 300 W
Blend 1 Nominal 250 W
Blend 2 Nominal 200 W
Blend 3 Nominal 150 W
Coag Nominal 120 W
Low Voltage Coag Nominal 99 W
Bipolar Nominal 70 W
Product Specification
Safety System
REMβ„’ Monitoring Continuous patient impedance monitoring
Safety Function Automatic shutdown on fault detection
Alarms Audible and visual alarms
Physical Specifications
Dimensions 13β€³ (W) Γ— 21β€³ (D) Γ— 8β€³ (H)
Weight 23 lbs
Cooling Convection cooling (no fan)
Leakage & Monitoring
Low Frequency Leakage <10 mA (normal), <100 mA (fault conditions)
High Frequency Leakage <150 mA RMS
REM Measurement Frequency 140 kHz
REM Measurement Current Max 3 mA
Resistance Range 5 – 135 ohms (REM pad)
Audio & Indicators
Indicator Tones Adjustable (minimum 45 dB at 1 meter)
Alarm Tones Fixed (minimum 65 dB at 1 meter)
Included / Optional Accessories
Monopolar Foot Switch Included
Bipolar Foot Switch Optional
Rolling Cart Optional